• CN
  • EN
  • Lubricating oil/detection and analysis category
    SC-430 portable oil pollution particle size analyzer
    This instrument adopts the light blocking counting principle specified by the International Hydraulic Standards Committee and is used as a rapid detection device for on-site oil contamination levels. It has the advantages of small size, light weight, fast
    • Details
    • Parameter
    • characteristic

             This instrument adopts the light blocking counting principle specified by the International Hydraulic Standards Committee and is used as a rapid detection device for on-site oil contamination levels. It has the advantages of small size, light weight, fast detection speed, high accuracy, and good repeatability, and can work under harsh conditions such as high temperature and high pressure. Built in micro water sensor and temperature sensor, which can detect both water content and oil temperature while detecting pollution level. Suitable for engine oil, gear oil, transformer oil (i.e. insulation oil), hydraulic oil, lubricating oil, synthetic oil and other fluids, it can be widely used in aerospace, petrochemicals, transportation ports, steel metallurgy, automobile manufacturing and other fields.

    1. Light source: semiconductor laser

    2. Flow rate range: 20-60mL/min

    3. Offline detection of sample viscosity: ≤ 100cSt, pressure chamber can be selected when viscosity is high

    4. Online pressure detection: 0.1-0.6Mpa (with optional pressure reducing device, the maximum pressure can reach 40Mpa)

    5. Particle size range: 1-500 μ m (using different types of sensors)

    6. Interface: USB interface, power interface

    7. Data storage: Provides 1000 sets of data storage space and supports USB storage

    8. Sensitivity: 1 μ m or 4 μ m (c)

    9. Limit coincidence error: 10000 particles/ml

    10. Counting volume: 1-999ml

    11. Counting accuracy: ± 0.5 pollution level

    12. Protection level: IP67

    13. Test interval: 1-24 hours

    14. Test sample temperature: 0-80 ℃

    15. Water activity reference value: 0-1aw (± 0.05aw)

    16. Water content: 0-120ppm (± 10%)

    17. Working temperature: -20-60 ℃

    18. Power supply: AC 220V ± 10%, 50/60Hz or DC12-40V

    19. Weight: 2.5kg

    20. Volume: 275 × 220 × 107mm


    1. Using the principle of photoresist (shading) method and high-precision laser sensor, it has small size, high accuracy, and stable performance

    2. Suitable for laboratory or on-site testing, and can also be equipped with pressure reducing devices for online high-pressure measurement, real-time monitoring of particle contamination in oil systems

    3. External pressure chamber can form positive/negative pressure to achieve detection and sample degassing of high viscosity samples

    4. Built in data analysis system, capable of displaying real data of particle size in each channel and automatically determining sample grade

    5. The pipeline is made of 316L and PTFE materials to meet the testing requirements of various organic solvents and oils

    6. Equipped with volume flushing and duration flushing modes, convenient for users to use and maintain the equipment

    7. Built in particle pollution level standards such as ISO4406, NAS1638, SAE4059, GJB420A, GJB420B, Г OCT17216, GB/T14039, etc

    8. Built in calibration function, can calibrate according to standards such as GB/T21540, ISO4402, GB/T18854, etc

    9. Built in data analysis system, capable of automatically determining sample grade based on standards, with automatic data processing and printing functions

    10. Can set any alarm level to achieve pollution or cleanliness detection

    11. Built in micro water sensor and temperature sensor

    12. Chinese and English input, one click switch, with preset, input, modification, and storage functions, easy and fast operation

    13. Ultra large storage, can be stored inside the instrument or on external storage devices

    14. Embedded design, high-strength casing, easy to carry, suitable for various types of construction machinery


    Related products

    Office Phone:+86 0731-85516578
    Fax:+86 0731-85535848
    Mobile phone:+86 139 7580 1381
    E-mail:cssc98@126.com
    Whatsapp/Wechat:24238498
    Address:Room 306, 3rd Floor, Building 4, No. 101, Hunan Provincial Research Achievement Transformation Center, Longping High tech Park, Furong District, Changsha City, Hunan Province

    Tiktok

    WeChat

    Video

    Mobile website

    Copyright (C) Changsha Sichen Instrument Technology Co., Ltd. 2024 All Rights Reserved

    在线客服系统

    MESSAGE

    Consultation Hotline:+86 400-660-9089
    Code