- Other/Petroleum Product Testing Category
This instrument is designed and manufactured according to the national petrochemical industry standard SH/T 0175-94 and with reference to the requirements of the American Society for Testing and Materials standard ASTM D2247-88. It is mainly used for the determination of oxidation stability (accelerated method) of distillate fuels. The instrument consists of an oxidation device, a dark box, and a filtering device, which can automatically control the oxidation temperature and oxidation time. Easy to operate and reliable in performance. This instrument is a metal bath and does not require water replenishment during the testing process. It is clean and pollution-free.
1. Power supply: AC220V ± 10% 50Hz
2. Main heating power: 1.6kw
3. Auxiliary heating power: 0.8kw
4. Constant temperature accuracy: ± 0.5 ℃
5. Number of test holes: 6 holes
6. Dimensions: Host 700 × 500 × 1480 (mm) Floor standing movable dark box 400 × 400 × 1100 (mm) Floor standing suction filter vacuum device 200 × 300 × 300 (mm) Desktop
7. Gas source: Oxygen (commonly used oxygen cylinders to be provided by customers)
1. This instrument has a desktop structure, simple and beautiful design, and meets the requirements of SH/T0175 standard
2. Embedded computer, industrial computer control, LCD display screen, fully Chinese operating system.
2. Equipped with a new gas pressure regulator, it can ensure that the output pressure remains constant even when the cylinder pressure drops.
3. There are 6 test holes equipped with 6 flow meters, which respectively regulate and indicate the flow rate of 6 oxidation tubes.
4. Equipped with a digital timer for setting and recording the time of the experiment.
5. Adopting advanced metal baths, it overcomes the problem of different boiling points caused by different atmospheric pressures, and has the advantages of fast heating, high accuracy, and no need for conversion.
6. The instrument uses solenoid valves to complete the oxygen filling and discharging of the oxygen bomb, making the operation process simple and easy. The solenoid valve connections are all made by hard connections, which maximizes the avoidance of system leakage.
7. During the experiment, the pressure curve and pressure drop curve are displayed. After the experiment is completed, the data can be saved
8. Close the oxygen shut-off valve directly after oxygenation to avoid system leakage affecting the test results.
9. The oxygen bomb adopts advanced hand rotation technology for sealing, which ensures good sealing performance inside the oxygen bomb without the need for special tools.
10. Equipped with an oxygen bomb installation overvoltage protection device.
11. Pressure gauge measuring range: 0-1.6MPa
12. The instrument is suitable for normal use in areas above 3650m altitude