    This instrument is designed and manufactured according to the requirements specified in the industry standard SH/T0325 "Determination of Oxidation Stability of Lubricating Greases" of the People's Republic of China. It is suitable for measuring the oxidation resistance of lubricating grease in static storage in an oxygen sealed system under elevated test temperature conditions. Method summary: Place the sample in an oxygen bomb heated to 99 ℃ and filled with 758kPa oxygen for oxidation. Observe and record pressure at regular intervals. After a specified time period, the degree of oxidation of the lubricating grease is determined by the corresponding decrease in oxygen pressure.

    1. Working power supply: AC220V ± 10% 50Hz

    2. Compliant with standard: SH/T0325 ASTM D942 IP142

    3. Temperature control range: room temperature~200 ℃

    4. Temperature control accuracy: ± 0.5 ℃

    5. Work unit: Two Bombs

    6. Temperature control method: microcontroller PID temperature controller

    7. Dish rack: made of alloy steel containing 18% chromium and 8% nickel

    8. Heating power: 1.8kw


    1. The instrument heating device adopts our special metal bath heating method, which heats uniformly and maintains a constant temperature. The advantages of no smoke pollution, no leakage, no maintenance, and constant temperature stability.

    2. The instrument is equipped with a new type of gas pressure regulator, which can ensure that the output pressure remains basically unchanged when the cylinder pressure drops.

    3. The instrument is equipped with two test baths, which can conduct tests on one sample or two samples simultaneously, making it convenient and flexible.

    4. Built in time and ultra-high voltage alarm system, safe and accurate.


    Related products

    Office Phone:+86 0731-85516578
    Fax:+86 0731-85535848
    Mobile phone:+86 139 7580 1381
    E-mail:cssc98@126.com
    Whatsapp/Wechat:24238498
    Address:Room 306, 3rd Floor, Building 4, No. 101, Hunan Provincial Research Achievement Transformation Center, Longping High tech Park, Furong District, Changsha City, Hunan Province

    Tiktok

    WeChat

    Video

    Mobile website

    Copyright (C) Changsha Sichen Instrument Technology Co., Ltd. 2024 All Rights Reserved

    在线客服系统

    MESSAGE

    Consultation Hotline:+86 400-660-9089
    Code