    SC-1792 Thiol Sulfur Analyzer
    This instrument is designed and manufactured in accordance with the requirements of the People's Republic of China standard GB/T 1792 "Determination of Thiol Sulfur in Distillate Fuels (Potentiometric Titration Method)", and is suitable for measuring thio
    • Details
    • Parameter
    • characteristic

         This instrument is designed and manufactured in accordance with the requirements of the People's Republic of China standard GB/T 1792 "Determination of Thiol Sulfur in Distillate Fuels (Potentiometric Titration Method)", and is suitable for measuring thiol sulfur in jet fuel, gasoline, kerosene, and light diesel fuel with a content ranging from 0.0003 to 0.01% (m/m) and without hydrogen sulfide. It is an ideal intelligent instrument for various petrochemical laboratories, analysis rooms, and research departments.

    1. Measurement range: 0.0003% to 0.01% (m/m)

    2. Accuracy: Relative error ≤ 5%

    3. Potential measurement range: 0 to ± 1999.5mv

    4. Basic error: 0.1% ± 0.5mv

    5. Relative error: ≤ 5%

    6. Burette volume: 10ml

    7. Accuracy of burette: ± 0.1% F · S

    8. Environmental temperature: 5 ℃~35 ℃

    9. Relative humidity: not exceeding 80%

    10. Working power supply: AC220V ± 10% 50HZ

    11. Instrument dimensions: 350mm × 280mm × 178mm

    12. Instrument weight (host): 10Kg;



    1. The Windows operating system adopts a human-machine dialogue mode for operation, which is simple and easy to understand.

    2. The key components of the titration device are all imported from Switzerland, with stable and reliable performance and extremely low transmission noise.

    3. The combination of automatic cleaning, automatic fluid replenishment, constant value liquid addition, and electromechanical organic technology.

    4. Multi parameter setting and correction make the analysis of different types of samples simpler, faster, with shorter analysis time and more accurate results.

    5. Automatic endpoint judgment, no indicator required, suitable for the analysis of thiol sulfur in various light and dark colored oils in petrochemical products. Titration results and data are automatically stored and printed, and complete titration data can be provided for analysis and research.

    6. Double high impedance input, three electrode system, more stable and reliable electrode potential.

    7. Supports RS232 connection to PC and comes with titration specific software as a bonus; Capable of data transmission;


    Related products

    Office Phone:+86 0731-85516578
    Fax:+86 0731-85535848
    Mobile phone:+86 139 7580 1381
    E-mail:cssc98@126.com
    Whatsapp/Wechat:24238498
    Address:Room 306, 3rd Floor, Building 4, No. 101, Hunan Provincial Research Achievement Transformation Center, Longping High tech Park, Furong District, Changsha City, Hunan Province

    Tiktok

    WeChat

    Video

    Mobile website

    Copyright (C) Changsha Sichen Instrument Technology Co., Ltd. 2024 All Rights Reserved

    在线客服系统

    MESSAGE

    Consultation Hotline:+86 400-660-9089
    Code